The
American Dental Association Council on JADA 2012;143(8):899-902 Scientific Affairs
The following resources related to this article are
available online at jada.ada.org ( this information is current as of November 2, 2012):
Updated
information and services including high-resolution figures, can be found in the online version of this article at: http://jada.ada.org/content/143/8/899
This article
cites 8 articles, 2 of which can be accessed free:
http://jada.ada.org/content/143/8/899/#BIBL
Information
about obtaining reprints of this article or about permission to
reproduce this article in whole or in part can be found at: http://www.ada.org/990.aspx
The use of
cone-beam computed tomography in
dentistry
An advisory statement from the
American Dental
Association
Council on Scientific Affairs
The American Dental Association Council on Scientific
Affairs
The emergence of
cone-beam computed tomography (CBCT) has expanded the field of oral and maxillofacial
radiology. CBCT imaging provides three-dimensional volumetric data construction
of dental and associated maxillofacial structures with isotropic resolution and
high dimensional accuracy.
A CBCT scanner uses a collimated
x-ray source that produces a cone- or pyramid-shaped beam of xradiation, which makes a single
full or partial circular revolution around the patient, producing a sequence of
discrete planar projection images using a digital detector. These
two-dimensional images are reconstructed into a three-dimensional volume that
can be viewed in a variety of ways, including cross-sectional images and volume
renderings of the oral anatomy.
Although CBCT units produce a
higher radiation dose than one would
receive from a single traditional dental radiograph, the radiation dose
delivered typically is less than that produced during a medical multichannel
computed tomographic scan. CBCT radiation doses also vary widely according to
the device used, x-ray energy and filtration, tolerance for image noise and motion
artifacts, and the size of the imaging area (field of view [FOV]) that is used
to acquire volumetric data 1-3 (Table,2,4,5 page 901).
Since CBCT devices were
introduced commercially in the United States
in 2001, dentists have come to use the technology in increasing numbers. Yet, although CBCT
technologies have advanced rapidly across time, concerns have been expressed
about whether the information acquired with CBCT imaging warrants the additional
exposure risk, as well as about the level of training, education and experience
required to interpret the CBCT data set.1
To provide guidance on CBCT
imaging, national and international groups have prepared basic principles,1
position
statements6,7and professional guidelines for
CBCT use.4,8 Recommendations for adequate
operator education regarding use
and interpretation of CBCT imaging also have been published.1,6
In
addition, CBCT guidance documents are being developedby dental specialty
organizations.9
As with any
clinical guidance regarding the acquisition of diagnostic information, dentists
must keep in mind their primary ethical obligation to protect patients from
harm. Consistent with its mission to serve as a primary resource on the science
of dentistry, the ADA Council on Scientific Affairs (CSA) reviewed the current
science, guidance and
other resources available from professional organizations to prepare this advisory
statement of principles for the safe use of CBCT in dentistry. The Council then
sought comments and input from a range of stakeholder organizations (listed in
the acknowledgments at the end of this article) to develop this collaborative guidance statement
for the profession.
PRINCIPLES FOR THE SAFE USE OF DENTAL AND MAXILLOFACIAL
CONEBEAM COMPUTED TOMOGRAPHY
The
Council recommends adherence to the following principles for the safe and
appropriate use of CBCT in clinical practice.
As with other radiographic
modalities, CBCT imaging should be used only after a review of the patient’s
health and imaging history and completion of a thorough clinical examination.
In accordance with the National
Council on Radiation Protection & Measurements’10 (NCRP’s) Report No. 145 and
standard selection criteria for dental radiographs,11 clinicians should perform
radiographic imaging, including CBCT, only after professional justification
that the potential clinical benefits will outweigh the risks associated with
exposure to ionizing radiation. All radiographic examinations should be
indicated clinically and justified appropriately, and such examinations should
not be performed for screening purposes. Additional considerations should be
weighed prior to the exposure of children and adolescents. These patients are
more radiosensitive (that is, their cancer risk per unit dose of ionizing
radiation is higher), and they have a longer lifetime risk of developing radiation-induced
cancers.12,13
The clinician should prescribe
traditional dental radiographs and CBCT scans only when he or she expects that
the diagnostic yield will benefit patient care, enhance patient safety,
significantly improve clinical outcomes or all of these.
CBCT should be considered as an
adjunct to standard oral imaging modalities. CBCT may supplement or replace
conventional (twodimensional or panoramic) dental radiography for the
diagnosis, monitoring and treatment of oral disease or the management of oral
conditions when, in the clinician’s decision-making process, he or she
determines that oral anatomical structures of interest may not be captured
adequately by means of conventional radiography.
In accordance with the
“as-low-as-reasonablyachievable” (ALARA) principle, radiation dose for dental
patients should be optimized to achieve the lowest practical level to address a
specific clinical situation. (Author’s
note: “Dose optimization” means
delivering a radiation dose to the organs and tissues of clinical interest no
greater than that required for adequate imaging and minimizing the dose to
other structures. The patient’s radiation dose is considered to be optimized
when imaging is performed with the least amount of radiation required to
provide adequate image quality. The goal of every imaging procedure is to
provide images adequate for the clinical purpose. What constitutes adequate
image quality depends on the modality being used and the clinical question
being asked.) The clinician should limit the radiation dose for CBCT scans by
optimizing image quality, using the smallest FOV necessary for imaging a
specific anatomical area of interest and using the lowest combination of tube
output and scan time (in milliamperes) consistent with adequate image noise
content and motion artifact.
CBCT operators should take every
precaution to reduce radiation dose and ensure the patient’s safety during CBCT
imaging. The use of thyroid collars and lead aprons is recommended in the
NCRP’s10
radiation safety guid-ance for the
dental profession. However, it is neither possible nor desirable to use these
protective devices in all clinical situations, especially in cases in which the
collar or apron may obstruct the area of interest. Protective thyroid collars
and lead aprons should be used when they will not interfere with the examination.
ABBREVIATION KEY. ADA:
American Dental Association. ALARA: As
low as reasonably achievable. CBCT:
Cone-beam computed tomography. CSA: Council
on Scientific Affairs. FOV: Field of view. NCRP:
National Council on Radiation
Protection & Measurements.
A CBCT examination should be
prescribed by a dentist who has appropriate training and education in CBCT
imaging, including an understanding of the significance of CBCT selection and
imaging findings.
CBCT images of the oral and
maxillofacial structures that are the subject of the CBCT examination should be
evaluated by a dentist with appropriate training and education in CBCT
interpretation. (Author’s note: Establishing formal standards for
CBCT training and education is beyond the scope of this CSA advisory statement.
The Council will share the statement with the Commission on Dental
Accreditation and other educational groups for further consideration.)
Regardless of the primary purpose
for the selection of CBCT, the complete image data set must be interpreted by
an appropriately qualified health care provider (such as a dentist or a
physician). The prescribing clinician should receive a thorough radiological
report. If the prescriber also interprets the CBCT images, he or she should
enter the findings into the patient record and communicate them appropriately
to the patient or, if the patient is a minor, to the patient’s parent or legal
guardian.
Dental practitioners who use CBCT
devices must receive appropriate training and education in the safe use of CBCT
imaging systems. Although there may be instances in which training provided by
vendors of CBCT systems is appropriate, dental practitioners should consider
the source of the information concerning radiation safety. The Council
encourages CBCT operators to participate in continuing education courses to
maintain adequate knowledge regarding radiation protection in the dental care
setting.
Dentists must abide by applicable
federal and state regulations in the provision of dental imaging modalities.
This includes following regulations or guidance to ensure a safe working
environment for both the staff and the public in relation to CBCT equipment and
other sources of ionizing radiation. CBCT unit operators should contact state
and local radiation control programs to verify any additional requirements for
operation of CBCT, including applicable requirements for licensure or
accreditation.
Dentists should use professional
judgment in the prescription and performance of CBCT examinations by consulting
recommendations from available CBCT guidelines and by considering the specific
clinical situation and needs of the individual patient. Given the ongoing
development and research in this technology, dentists should stay abreast of
the scientific literature and apply an evidence- and science-based approach to
the use of CBCT.
This advisory statement calls for
appropriate agencies within the ADA and the dental community at large to
develop and implement recommendations and criteria for adequate CBCT training
and education of dentists or other CBCT unit operators. These recommendations
should include but not be limited to patient evaluation, radiation protection,
selection of appropriate CBCT imaging parameters, performance of the CBCT
examination and image interpretation. The recommendations also must include
requirements for predoctoral dental education programs and for continuing
education coursework and training.
Facilities considering the
installation of CBCT devices should consult a health physicist (or other
qualified expert) to perform a shielding analysis based on NCRP reports 145 and
147.10,14
Facilities using CBCT systems
should consult a health physicist (or other qualified expert) to perform
equipment performance and compliance evaluations initially at installation and
then follow a schedule in compliance with local, state and federal
requirements. The Council recommends that a performance evaluation be completed
at least annually. The evaluations should include patient dose estimation to
assist the facility with patient dose management.
Staffs of facilities using CBCT
should establish a quality control program. This program can be based on the
manufacturer’s recommendations or can be established, implemented and monitored
by a qualified expert.
SUMMARY
CBCT
technologies offer an advanced point-ofcare imaging modality that clinicians
should use selectively as an adjunct to conventional dental radiography. The
selection of CBCT for dental and maxillofacial imaging should be based on
professional judgment in accordance with the best available scientific
evidence, weighing potential patient benefits against the risks associated with
the level of radiation dose. Clinicians must apply the ALARA principle in
protecting patients and staff during the acquisition of CBCT images. This includes
appropriate justification of CBCT use, optimizing technical factors, using the
smallest FOV necessary for diagnostic purposes and using appropriate personal
protective shielding.
Disclosure. Members
of the American Dental Association Council on Scientific Affairs (CSA) are
required to maintain a current conflict-of-interest disclosure to participate
in CSA activities. No potential conflicts of interest relevant to this
statement were reported.
The
American Dental Association (ADA) Council on Scientific Affairs gratefully
acknowledges Dr. John Ludlow, professor, School of Dentistry, University of
North Carolina, Chapel Hill, and Dr. Sharon Brooks, professor emerita, School
of Dentistry, University of Michigan, Ann Arbor, for their assistance in
finalizing this advisory statement. The ADA expresses its appreciation to the
following organizations for their significant contributions to the statement:
the American Academy of Oral and Maxillofacial Pathology; the American Academy
of Oral and Maxillofacial Radiology; the American Academy of Pediatric
Dentistry; the American Academy of Periodontology; the American Association of
Endodontists; the American Association of Oral and Maxillofacial Surgeons; the
American Association of Orthodontists; the American Association of Physicists
in Medicine; the Conference of Radiation Control Program Directors; the
National Council on Radiation Protection & Measurements; and the U.S. Food
and Drug Administration.
1. Horner K, Islam M, Flygare L, Tsiklakis K, Whaites
EJ. Basic principles for use of dental cone beam computed tomography: consensus
guidelines of the European Academy of Dental and Maxillofacial Radiology.
Dentomaxillofac Radiol 2009;38(4):187-195.
2. Pauwels R, Beinsberger J, Collaert B, et al;
SEDENTEXCT Project Consortium. Effective dose range for dental cone beam
computed tomography scanners. Eur J Radiol 2012;81(2):267-271.
3. Ludlow JB, Ivanovic M. Comparative dosimetry of
dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral
Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106(1):106-114.
4. The SEDENTEXCT Project. Radiation Protection: Cone
Beam CT for Dental and Maxillofacial Radiology: Evidence Based Guidelines 2011
(v2.0 Final). www.sedentexct.eu/files/guidelines_final.pdf. Accessed May 11,
2012.
5. Ludlow JB, Davies-Ludlow LE, White SC. Patient risk
related to common dental radiographic examinations: the impact of 2007
International Commission on Radiological Protection recommendations regarding
dose calculation. JADA 2008;139(9):1237-1243.
6. Carter L, Farman AG, Geist J, et al; American
Academy of Oral and Maxillofacial Radiology. American Academy of Oral and
Maxillofacial Radiology executive opinion statement on performing and
interpreting diagnostic cone beam computed tomography. Oral Surg Oral Med Oral
Pathol Oral Radiol Endod 2008;106(4):561-562.
7. Joint Position Statement of the American Association
of Endodontists and the American Academy of Oral and Maxillofacial Radiology.
Use of cone-beam computed tomography in endodontics.
www.aaomr.org/resource/resmgr/Docs/AAOMR-AAE_postition_ paper_CB.pdf. Accessed
July 10, 2012.
8. Academy of Osseointegration. 2010 guidelines of the
Academy of Osseointegration for the provision of dental implants and associated
patient care. Int J Oral Maxillofac Implants 2010;25(3):620-627.
9. Scarfe WC. “All that glitters is not gold”:
standards for conebeam computerized tomographic imaging (published online ahead
of print Feb. 3, 2011). Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;111(4):402-408.
doi:10.1016/j.tripleo.2011.01.006.
10. National Council on Radiation Protection &
Measurements. Radiation Protection in Dentistry (Report No. 145). Bethesda,
Md.: NRCP Publications; 2003.
11. American Dental Association; U.S. Department of Health
and Human Services. The Selection of Patients for Dental Radiographic
Examinations. Chicago: American Dental Association; 2004.
www.ada.org/sections/professionalResources/pdfs/topics_radiography_examinations.pdf.
Accessed May 11, 2012.
12. National Research Council of The National
Academies, Committee to Assess Health Risks from Exposure to Low Levels of
Ionizing Radiation, Board of Radiation Effects Research, Division on Earth and
Life Studies. 2006. Health Risks From Exposure to Low Levels of Ionizing
Radiation: BEIR VII phase 2. Washington: National Academy of Sciences, National
Academies Press; 2006.
13. International Commission on Radiological
Protection. The 2007 Recommendations of the International Commission on
Radiological Protection (publication 103). Ann ICRP 2007;37(2-4):1-332.
14. National Council on Radiation Protection &
Measurements. Structural Shielding
Design for Medical Imaging Facilities (Report No. 147). Bethesda, Md.: NRCP
Publications; 2004.
Niciun comentariu:
Trimiteți un comentariu